HPLC dilution and injection calculator
Three calculations that decide whether an injection is informative: how much stock goes into the flask, how much peptide reaches the column, and what the dilution series actually contains. Arithmetic only — no protocol, no method recommendation.
1. Stock to working concentration
Defaults are illustrative. Replace them with your own values; nothing is stored.
Dilution factor: —
Stock to pipette: —
Diluent to add: —
2. Mass injected on column
Mass on column: —
Amount in moles: —
3. Serial dilution series
Transfer volume per step: —
Concentrations: —
The arithmetic
- Dilution: C₁V₁ = C₂V₂, so stock volume = (target concentration × final volume) ÷ stock concentration. Diluent volume = final volume − stock volume.
- Injection mass: concentration in µg/mL equals ng/µL, so mass on column (ng) = concentration (µg/mL) × injection volume (µL).
- Moles: pmol = ng × 1000 ÷ molecular weight in g/mol.
- Serial dilution: each tube is the previous concentration divided by the dilution factor; with a fixed volume per tube, the volume transferred at each step is tube volume ÷ dilution factor.
Worked with the defaults: 2 mg/mL = 2000 µg/mL; 100 × 10 ÷ 2000 gives 0.5 mL (500 µL) of stock in a 10 mL flask, a 20-fold dilution. At 100 µg/mL and a 10 µL injection, 1,000 ng reaches the column. A 1,000 µg/mL top standard diluted 2-fold over six points gives 1000, 500, 250, 125, 62.5 and 31.25 µg/mL, transferring 0.5 mL into 0.5 mL of diluent at each step.
What the calculator does not know
- Whether the stock concentration is real. A nominal concentration computed from fill weight ignores peptide content, counter-ion and water. The inputs must come from a measured or corrected value — see net peptide content.
- Column capacity and void volume. 1,000 ng on a 2.1 mm column is a very different load from 1,000 ng on a 4.6 mm column. The number is a mass, not a suitability judgement.
- Sample solvent strength. Injecting in a solvent stronger than the mobile phase distorts early peaks regardless of how accurately the dilution was made.
- Adsorption losses. At low concentrations, a meaningful fraction of peptide can bind to the flask, the vial or the needle — see adsorption to labware.
- Detector linearity. The top standard must sit inside the validated range of the method, not merely inside the detector's nominal maximum.
- Volumetric glassware accuracy. A 10 mL flask and a 100 µL pipette have tolerances; at low volumes they dominate the error budget.
Practical notes on making the dilution
- Bring the stock and the diluent to the same temperature before pipetting; thermal expansion of a small volume is not negligible relative to a 100 µL transfer.
- Rinse the pipette tip with the stock before transferring, especially for a solution containing organic solvent.
- Add the stock into the diluent, not the reverse, and mix by inversion rather than vortexing if the peptide is prone to aggregation.
- Prepare the working solution fresh where the method allows it; a dilute solution has more surface area per unit mass and less stability than the stock.
- Record the actual values — stock lot, measured concentration, pipette volumes, diluent composition — so that a later result can be reconstructed.
Related reading & products
- Reading an HPLC chromatogram — peak shape, integration and area percent
- Method validation — why the range and LOQ matter before the injection
- Measuring concentration — verifying the stock before diluting it
- Adsorption to labware — the low-concentration loss
- Reconstitution calculator — vial to stock solution
- All tools
- All knowledge guides